top of page
Search

A 3D-Printed Patient-Specific Cutting Guide for Severe Post-Traumatic Genu Valgum in a Child: Case Report

turannajafov
Oct 4
4 min read

Updated: Oct 19

Line drawing of a 3D-printed femur model after guided closing-wedge osteotomy, with the removed bone wedge
Line drawing of the 3D-printed femur model after guided wedge resection: planned osteotomy line and the removed bone wedge.

Patient-specific cutting guides are widely used in adult arthroplasty and corrective osteotomy, yet reports in the skeletally immature femur remain scarce. This is how we used a surgeon-designed, 3D-printed guide to correct a 30° post-traumatic genu valgum in a 9-year-old boy: from CT to a rehearsed osteotomy on a printed model, and from there to a sterile guide in theatre.

Key points

  • 30° distal femoral valgus after an untreated fracture of the lateral distal femoral physis

  • Two cutting slots set at 30° on a guide body built from a 5-mm offset of the bone surface

  • Mock osteotomy on a PLA femur with the actual operating-room saw before sterile production

  • Guide placement to wedge removal: under 30 minutes

  • Planned 30° correction achieved; weight-bearing at two months and a normalised gait

The case

A 9-year-old boy had sustained a crush injury four years earlier, including a circumferential degloving of the right thigh that was treated with skin grafting. An associated fracture involving the lateral distal femoral physis was not treated at the time. Lateral growth arrest followed, and the knee drifted into progressive valgus.

At presentation:

  • 30° genu valgum on full-length standing radiographs, with the CORA 1 cm proximal to the physis

  • MCL insufficiency

  • Equinus contracture

  • Knee flexion limited to 90°

Pre-operative full-length radiograph of post-traumatic genu valgum and post-operative AP femur after guided varus osteotomy with plate fixation
Left: pre-operative standing full-length radiograph showing the right-sided genu valgum. Right: post-operative day 1 AP femur: planned correction, fixed with a medial anatomical distal femoral plate and a retrograde cannulated screw.

Results of any surgical or interventional procedure may vary from person to person. Please consult your physician for detailed advice before any procedure.

Why a guide in this case?

The deformity was large and sat in a short metaphyseal segment just above the physis, so there was little margin for error in wedge size or hinge position. The soft-tissue envelope was grafted and scarred, which made a predictable, rehearsed procedure with no repeated cuts especially attractive.

Planning

  1. Segmentation: CT data were segmented and remeshed in medical imaging software.

  2. Deformity analysis: The CORA was located 1 cm proximal to the physis. A medial closing-wedge varus osteotomy was planned at the CORA, hinging on the lateral cortex.

  3. Guide design: Two cutting blocks were set at 30° to each other on a guide body created by a 5-mm offset of the bone-contact surface, with 2.5-mm saw-blade slots and pin holes for fixation.

  4. Rehearsal: The femur and the guide were first printed in PLA. A mock osteotomy with the actual operating-room saw confirmed fit, seating and blade trajectory.

  5. Final guide: The guide was then printed in biocompatible resin and sterilised.

Planning schematic of a medial closing-wedge varus osteotomy at the CORA with a patient-specific cutting guide
Planning schematic: CORA 1 cm proximal to the physis, medial closing wedge of 30° with a lateral hinge, and a pinned patient-specific guide with two cutting slots.
Mock osteotomy on a 3D-printed PLA femur with the operating-room saw through a patient-specific cutting guide
Rehearsal on the 3D-printed femur: the operating-room saw through the guide slots (left), and the model after wedge resection (right).

Surgery

Through an anteromedial approach, extended distally and anteriorly for the MCL plication, the guide seated on the medial distal femur as rehearsed and was pinned. The saw followed the two slots, and the 30° medial wedge was removed at the CORA. Guide placement to wedge removal took under 30 minutes.

Sterile 3D-printed resin cutting guide pinned on the medial distal femur during surgery
Sterile resin guide pinned on the medial distal femur.
Intra-operative fluoroscopy showing the two guide-defined osteotomy cuts of the closing wedge
Intra-operative fluoroscopy: the two guide-defined cuts outlining the closing wedge.

In the same session:

  • Medial and lateral physeal ablation to prevent recurrence from the asymmetric physis

  • Fixation with a medial anatomical distal femoral plate and a retrograde 5-mm cannulated screw

  • MCL plication

Results

  • The planned 30° correction was achieved with stable fixation.

  • No wound complication or infection occurred, despite the grafted soft-tissue envelope.

  • The equinus contracture was treated with a staged gastrocnemius recession at six weeks.

  • Knee motion started at one month and weight-bearing at two months; gait normalised.

Because the distal femoral physis was ablated at nine years of age, limb length and alignment are followed until skeletal maturity.

Take-home points

  • Rehearse before you sterilise. A printed model and the real saw verify fit, trajectory and wedge size before the final guide is produced.

  • The plan does the thinking. In theatre, the work came down to seating, pinning and cutting.

  • It combines well. A guided osteotomy can be performed together with same-session epiphysiodesis and soft-tissue procedures.

  • Know the limits. This is a single case. A guide needs CT, design and printing lead time, and an exposure that allows the guide to seat on bone. It does not replace deformity analysis or surgical judgement.

Disclosure and consent

Conflict of interest: The patient-specific cutting guide was designed by the author and manufactured by SPINEAST, a company owned by the author. No funding was received for this work.

Consent: Written informed consent for publication was obtained from the parents. Identifying details have been removed.

Questions, similar cases or collaboration: info@spineast.com · Türkçe versiyon: Bacak Eğriliği Ameliyatında Kişiye Özel 3D Kesi Rehberi · Patients: book an online appointment

This article is for professional information and does not replace an individual medical consultation.

Author and editor: Op. Dr. Turan Najafov, Orthopaedic and Traumatology Surgeon, specialised in Microsurgery, Deformity and Tumour Surgery · info@spineast.com · Last updated: October 2026

 
 
 

Comments


Commenting on this post isn't available anymore. Contact the site owner for more info.